Multifunction Wireless Phase Voltmeter

Send Inquiry
Multifunction Wireless Phase Voltmeter
Details
MOEORW-WNS42 new generation of intelligent three-phase AC power parameter measuring instrument, dual-unit wireless model, accurately detects power line parameters and wiring status.
Wireless connectivity, multi-parameter measurement, intelligent wiring identification, supports Bluetooth/PC data management, and dual power supply for safe operation.
Category
Electric Energy Test Equipment
Share to
Description

Multifunction Wireless Phase Voltmeter

 

Product Introduction

 

   This portable wireless phase voltmeter is engineered for on-site three-phase power inspection, wiring verification and power grid parameter measurement. It supports wireless synchronous data transmission between dual units, enabling non-contact, safe and efficient testing for voltage, current, phase angle, frequency and power factor. Lightweight and easy to carry, it adapts to complex outdoor and substation working environments, effectively simplifying field power detection workflows.

 

Product Parameter (specification)

 

1.Reference working conditions

Impact

Reference value or range

Allowable deviation

Ambient temperature

23℃

±5℃

Ambient humidity

(45~75)%RH

-

working power supply

Built-in lithium battery power

-

External electromagnetic interference

Should be avoided

-

Work location

The conductor being measured is positioned at the center of the clamp jaws.

-

Waveform of the measured signal

Sine wave (distortion factor β), see note 1

β=0.01

Frequency of the signal under test

50 Hz

±2 Hz

Phase frequency measurement

Range of amplitude of the measured signal

Voltage: 100V±25V

Current: 1A±0.2A

-

2. Basic Error Limit

Measurement items

range

Measurement range

resolution

Basic error

AC voltage

100V/ 50 0V

5V ~ 500V​​

0.001V

± 0.2 % of the range

Alternating current

200mA / 1A / 5A

5 mA ~ 10 A

0.1 mA

±( 0.2 % of the range + 0.3 % of the reading)

Phase measurement

0-360°

200 mA - 10A​

0.1°

± 1.0°

5mA ~ 200mA​

0.1°

± 2.0°

Frequency measurement

50 Hz

45-65Hz

0.01 Hz

± 0.02 Hz

Active power

(PF=1.0)

COS Φ ≥0.5L(C)

0.1W

± 0.5 % of the range

reactive power

(PF=1.0)

SIN Φ ≥0.5L(C)

0.1Var

±1.0% range

working power supply

Main unit: Powered by two 2200mAh/ 7.4V lithium batteries, via PT terminal power supply, and via adapter power supply .

Slave unit: Powered by 2000 mAh/ 3.7V lithium battery or adapter.

Charging time > 8 hours, charging cycles > 500 times

Note: For phase measurements at 5mA-10mA and 5V-10V , the phase error is ±3°. Under rated operating conditions, the rated operating error limit of each measurand shall not exceed twice the corresponding basic error limit.

3. Measurement Characteristics Description

Voltage channel

Input method: Three inputs are connected via UAN/UBN/UCN respectively; Input impedance: 500KΩ;

Measurement method: True RMS principle.

Current channel

Input method: The three inputs are sampled by clamp-on current transformers IA, IB, and IC respectively;

Measurement method: True RMS principle.

Online Records

Set the following items: U1, U2, U3, I1, I2, I3, Φ, P1, P2, P3, P, Q1, Q2, Q3, Q, cos1, cos2, cos3, cos, F (choose any 4 items each time).

Periodic interval: can be set arbitrarily from 1 to 99 seconds. Start and end times: Manual start

4. Safety Features

Withstand voltage: The voltage input terminal, the AC charging power input terminal and the instrument housing, and the iron core and lead wires of the clamp current transformer and the clamp handle can withstand a 1KV/50Hz sinusoidal AC voltage for 1 minute.

Insulation resistance: The insulation resistance between the test points mentioned above is greater than 100MΩ/1KV.

5. Instrument dimensions and weight

host Dimensions: 275×200×72 (mm) Weight: 1.2 kg (including battery)

from Machine dimensions: 185×85×41 (mm) Weight: 350 g (including battery)

Current clamp: 1 40 × 4 0 × 19 (mm) Weight: 160g/ piece; Jaw: Φ8mm

machine Box (main unit): 423 × 336 × 126 (mm) Weight: 5.6 kg

machine Box (slave unit): 280 × 240 × 127 (mm) Weight: 0.85 kg

6. Usage conditions

Operating environment: Temperature -15℃ to +40℃; Operating humidity ≤90%RH

Storage environment: Temperature -30℃ to +60℃; Storage humidity ≤95%RH

Altitude: <1500 meters

 

Product features and applications

 

  • Dual-host wireless linkage design, long transmission distance and strong anti-interference ability
  • High-precision sampling chip, ensuring stable and accurate measurement data
  • Integrated multiple measurement functions, one device for multi-index testing
  • Clear LCD display with intuitive data reading, support for real-time data refresh
  • Rechargeable built-in battery, long endurance for all-day field operation
  • Compact body with dustproof and drop-resistant structure, suitable for harsh working conditions

 

Prodection detail

 

    Widely applicable to power distribution line inspection, substation daily maintenance, industrial enterprise power consumption detection, electrical wiring phase sequence verification and power equipment routine troubleshooting.

 

Company Qualifications

 

 

FAQ

1.Is wireless connection stable in strong electromagnetic environment?

Built-in anti-interference module, which can maintain stable signal connection in substation and high-voltage electromagnetic scenes.

2.What is the maximum wireless transmission distance?

The effective communication distance can reach more than 100 meters in open space, meeting most on-site test demands.

3.Does the device support outdoor long-time use?

Equipped with large-capacity lithium battery, it realizes long-hour continuous work, with low power consumption design.

 

Hot Tags: multifunction wireless phase voltmeter, China multifunction wireless phase voltmeter manufacturers, factory, Double Clamps Digital Phase Meter, Three phase clamp on phase voltmeter, Power Quality Analyzer, Meter Calibration Test Bench, Energy Meter Field Calibrator, Three Phase Energy Meter Test Bench

Send Inquiry