Product Introduction
MOEORW-WBJ83 is a high-precision fully automatic tester designed for liquid insulation materials performance testing. Adopting advanced digital measurement technology and three-electrode oil cup structure, it effectively eliminates interference from stray capacitance and leakage current. The instrument supports full-automatic operation including heating, temperature control, data acquisition, calculation, display, printing and storage, with stable performance and high reliability, widely used in power equipment preventive test and insulation oil quality inspection.
Product Parameter (specification)
|
Test voltage |
0~2000V AC, 0~500V DC |
|
Temperature range |
room temperature~120℃ |
|
Dielectric loss factor |
0.00001~100, accuracy ±(0.5% of reading+0.0001) |
|
DC resistivity |
2.5MΩ·m~100TΩ·m, accuracy ±10% of reading |
|
Relative permittivity |
0~40, accuracy ±(0.5% of reading+0.1) |
|
Capacitance |
50pF~200pF, accuracy ±(0.5% of reading+0.1pF) |
|
Resolution |
tanδ 0.00001; resistivity 0.001MΩ·m; capacitance 0.01pF |
|
Power |
500W |
|
Dimensions |
470mm×425mm×385mm |
|
Weight |
22kg |
|
Working environment |
0℃~40℃, humidity ≤75%RH |
Product Feature
- Standard three-electrode oil cup (2mm gap), effectively suppress interference and improve test accuracy.
- Medium-frequency induction non-contact heating, fast speed, uniform temperature and precise control.
- Large color touch screen, intuitive Chinese menu, easy human-computer interaction.
- Built-in high-stability standard capacitor, ensure long-term measurement accuracy.
- Automatic calibration for empty cup, calibration data auto-saved for precise calculation.
- Multiple safety protections: cover open high-voltage cut-off, electrode short-circuit alarm.
- Auto store 100 groups test data with real-time clock, fast thermal printing.
Product Application
- Transformer, breaker, capacitor and other power equipment insulation oil preventive test.
- Insulation oil delivery inspection, factory inspection and performance verification.
- Laboratory research and quality control of liquid insulating materials.
- Power supply bureau, power station, transformer manufacturer, research institute and testing institution.
Company Qualifications
FAQ
Q1: Why no temperature signal detected during heating?
A: Check if the temperature signal cable is properly inserted into the oil cup temperature hole.
Q2: How to handle abnormal empty cup capacitance or tanδ?
A: Re-clean the oil cup as standard procedure; check signal line connection and RF connector.
Q3: What is the standard empty cup reference value?
A: Capacitance 60pF±5pF, tanδ ≤0.00010; unqualified means need re-cleaning.
Q4: How to properly clean the electrode cup for precision test?
A: Use petroleum ether, acetone, neutral detergent and distilled water for step cleaning, oven dry at 105~110℃ for ≥1 hour, assemble with silk gloves.
Q5: The instrument does not heat up, what to check?
A: Check if the heating fuse is blown; confirm no abnormality in the induction heating module.
Hot Tags: insulation oil dielectric loss & resistivity tester high precision, China insulation oil dielectric loss & resistivity tester high precision manufacturers, factory, Insulating Oil Dielectric Loss And Resistivity Tester, Insulating Oil Withstand Voltage Tester, Single cup Insulating Oil Withstand Voltage Tester, Insulation Oil Break Down Test Set, transformer oil bdv tester, Transformer Tan Delta Tester